Square and rectangular locator cones

Square and rectangular locator cones

Generate PDF
 
Square and rectangular locator cones for performing X-rays while reducing patient and practitioner dose.

The locator cone allows a reduction of the patient dose thanks to the forced collimation of the beam and the elimination of the secondary radiation at the exit of the tube.
The reduction of the patient dose leads to a reduction of the scattered radiation and therefore the dose to the practitioner.

It allows the restriction of the field covered on the equipment equipped with an automatic bucky.

The locator cone is fixed under the diaphragm in the rails provided for this purpose.

When ordering, specify the center distance of the rails to deliver the locator cone with the correct base width.
As standard the cone is delivered with a base of 176x176mm in stainless steel thickness 2mm.

Designation Code SSD (m) Covered field (mm)
240x300: 2 18A 1.10 m 120 x 150 mm
Stomach 11A 1.10 m 150 x 200 mm
Skull 9A 1.10 m 180 x 240 mm
Skull 12A 1.10 m 240 x 300 mm
Skull column 7A 1.10 m 150 x 400 mm
Column 15A 1.10 m 200 x 400 mm
Digestive 10A 1.10 m 300 x 400 mm
Lungs 16A 1.80 m 350 x 350 mm
Gross weight
3,5
Net weight
2,28